Here is a few simple steps to follow.
Homepage → Rack requisition → Start Entry → Key in Sample ID → Profile ( select accordingly) → Demography ( Fill up the rest of the patient details) → Entry → Play → Start
Before I load the sample to the rack, I had undergo the sample checking. And I found out that the serum is too little which might cause the probe of the analyser come into contact with the gel. Therefore I used an alternative method which shown by the photo below.
Low volume of serum is poured into a sample cup.
Besides, today was so lucky that we managed to find some fibrin exist in the sample. The fibrin was so troublesome and had prolonged the Turn around time. We tried to rerun the test for three times every after the noticeable fibrin was removed.
Detected Fibrin
Method to remove : use a toothpick to swirl it out.
Once the fibrin is not totally removed, the result will not be accurate or will not be completed.
Factors that affect the accuracy of analysis result :
- Sample integrity ( Haemolyse, Jaundice, Lipidemic ......)
- Separating Gel (Insufficient serum) / Sample clot (Fibrin)
- Reagent stability